beautypg.com

Power-on and remote operation default settings – AMETEK XEL Series 60V and below User Manual

Page 37

background image


Power-on and Remote Operation Default Settings

The following instrument status values are set at power on:

Status Byte Register

= 0

Service Request Enable Register †

= 0

Standard Event Status Register

= 128 (pon bit set)

Standard Event Status Enable Register †

= 0

Execution Error Register

= 0

Query Error Register

= 0

Parallel Poll Enable Register †

= 0

Registers marked thus are specific to the GPIB section of the instrument and are of limited use

via other interfaces.

The instrument will be in local state with the front panel controls active (providing rear panel
switches CV and CC are set to Local.

The instrument parameters at power-on are the same as at last switch off with the exception of
the output status. By default this is always off at power on but the user may change this from the
front panel to the same at power on as at switch off, see Switching On and Power-on Conditions
paragraph in the Manual Operation section.

The *RST (reset) interface command resets the instrument to the Remote Operation Default
settings.

Remote Operation Default settings are:
V

out

=0.1V, I

out

=100mA, DeltaV=10mV, DeltaI=1mA, Vmin=3V, Vmax=5V, Tracking Ratio = 100%,

Lock cancelled, V-Span cancelled, 500mA Range cancelled, Meter Average cancelled, OVP &
OCP at 5% above instrument range maximums.

LAn Err

message (10 seconds) for no LAN

connection found at power on is enabled. Configure page password on web server is reset to no
password.

Remote interface settings, stored set-ups, Vmin/Vmax values and Output state at power-on
setting are unchanged.

Mode and Sense settings are as determined by the front panel switches.

33

This manual is related to the following products: