IO?

Returns the output readback current for output
The response is A where is in Amps

IRANGE

Sets the current range of output to , where is 1 for Low (500mA)
range, 2 for High range. Output must be switched off before changing range.

IRANGE?

Report the current range for output .
The response is , where is 1 for Low (500mA) range, 2 for
High range.

DELTAV

Set the output voltage step size to Volts

DELTAI

Set the output current step size to Amps

DELTAV?

Returns the output voltage step size
The response is DELTAV where is in Volts.

DELTAI?

Returns the output current step size
The response is DELTAI where is in Amps.

INCV

Increment the output voltage by the step size set for output

INCVV

Increment the output voltage by the step size set for output and verify.

DECV

Decrement the output voltage by the step size set for output

DECVV

Decrement the output voltage by the step size set for output and verify

INCI

Increment the output current limit by the step size set for output

DECI

decrement the output current limit by the step size set for output

OP

set output on/off where has the following meaning: 0=OFF, 1=ON

OPALL

Simultaneously sets all outputs on/off where has the following meaning:
0=ALL OFF, 1=ALL ON.

If OPALL sets all outputs ON then any that were already on will remain ON
If OPALL sets all outputs OFF then any that were already off will remain OFF

OP?

Returns output on/off status.
The response is where 1 = ON, 0 = OFF.

TRIPRST

Attempt to clear all trip conditions.

LOCAL

Go to local. This does not release any active interface lock so that the lock
remains with the selected interface when the next remote command is received.

IFLOCK

Request interface lock. This command requests exclusive access control of the
instrument. The response is 1 if successful or –1 if the lock is unavailable either
because it is already in use or the user has disabled this interface from taking
control using the web interface

IFLOCK?

Query the status of the interface lock. The return value is 1 if the lock is owned
by the requesting interfaced instance; 0 if there is no active lock or –1 if the lock
is unavailable either because it is already in use, or the user has disabled this
interface from taking control using the web interface.

IFUNLOCK

Release the lock if possible. This command returns the value 0 if successful.
If this command is unsuccessful –1 is returned, 200 is placed in the Execution
Register and bit 4 of the Event Status Register is set indicating that there is no
authority to release the lock.

LSR?

Query and clear Limit Event Status Register .
The response is . See Status Reporting section for details.