Modbus register mapping – SOR 815DT Smart User Manual
Page 18
18/24
Form 1588 (09.14) ©SOR Inc.
Modbus
Register
Register Definition
Read/Write
"Data
Type"
Description
0
Loop current/voltage
R
Float
Current (mA)/Voltage (V)
4
PV unit
R
UINT8
PV units code
8
PV
R
Float
PV
12
SV unit
R
UINT8
SV units code
16
SV
R
Float
SV
20
TV unit
R
UINT8
TV units code
24
TV
R
Float
TV
28
FV unit
R
UINT8
FV units code
32
FV
R
Float
FV
36
Current and percent of range
R
Float
Percentage of range
40
Modbus slave ID
R/W
UINT8
Modbus Slave ID
44
Device Status
R
UINT32
Unit status flags
48
R
Float
Temperature
52
R
Float
AD5421 loop voltage
56
R
Float
VO_S voltage
60
R
Float
VREGOUT_S voltage
64
Device Information
R
UINT16
Main product ID
68
R
UINT16
Main model ID
72
R
UINT32
Main serial number
76
R
UINT8
BSL firmware build
Modbus Register Mapping
HART
Command
Description
Command Data
Command Reply Data
Byte
Description
Byte
Description
200
Write
password
0-6
Password
0-6
Access level
201
Enable/
Disable
write
protection
Read only
0
Write protect
code
202
Change
user
password
0-6
New password
0-6
New
password
203
Change
service
password
0-6
New password
0-6
New
password
204
Return
to default
passwords
None
Write only
The following table summarizes the Modbus register mapping assignments for the 815DT.