Hanna Instruments HI 3512 User Manual
Page 69

69
• For each record:
• The number of the record (3 chars)
• The reading status (1 char): R, O, U
• The unit of the reading (2 chars): it is sent only
for ISE, EC, Resistivity, TDS, Salinity log only
• The calculated reading, with sign, decimal point
and exponent (12 chars)
• The temperature reading, with sign and decimal
point (7 chars)
• The second reading status (1 char): R, O, U; not
available for EC log
• The second unit of the reading (2 chars); it is sent
only for Resistivity, TDS and Salinity log only
• The second reading, with sign and decimal point
(8 chars); not available for EC log
• The logging time (6 chars): it is sent only for
AutoEnd log
• The temperature probe persence (1 char)
LODxPyyy Requests the yyy
th
pH record logged data when x=1
LODxMyyy Requests the yyy
th
mV/Rel mV record logged data when x=1
LODxIyyy Requests the yyy
th
ISE record logged data when x=1
LODxEyyy Requests the yyy
th
EC record logged data when x=2
LODxRyyy Requests the yyy
th
Resistivity record logged data when x=2
LODxTyyy Requests the yyy
th
TDS record logged data when x=2
LODxNyyy Requests the yyy
th
Salinity record logged data when x=2
LODxPALL Requests all pH Log on demand when x=1
LODxMALL Requests all mV/Rel mV Log on demand when x=1
LODxIALL Requests all ISE Log on demand when x=1
LODxEALL Requests all EC Log on demand when x=2
LODxRALL Requests all Resistivity Log on demand when x=2
LODxTALL Requests all TDS Log on demand when x=2
LODxNALL Requests all Salinity Log on demand when x=2
The answer string contains:
• The number of records (3 chars): this is sent only is
“ALL” data is requested
• The log type (2 chars)
• The number of the record (3 chars)