The relative mV value calibration may be used to perform a single point
calibration with an ORP standard or remove the contribution of the
reference electrode to display mV equivalent to a SHE.
• Press

CAL when the instrument is in Relative mV measurement

mode. The relative mV value and the temperature values are
displayed.

• Use the

ARROW keys if you want to change the displayed relative

mV value.

• Press

Zero function key if you want Rel mV reading to be zero

(Relative mV offset equals the mV reading).

• When the reading is stable in mV range and the Relative mV offset

is inside the offset window (±2000 mV), the

Confirm function key

is displayed

.

• Press

Confirm to confirm relative mV calibration. The instrument will

return to measurement mode.

• If the absolute mV reading is out of range or the Relative mV offset

is out of the offset window,

“Wrong relative offset” message is

displayed.

Change the input value or the Relative mV value to complete the
calibration process.

Note: If a Rel mV offset calibration exists, CLR function key is

displayed. Press

CLR if you want Rel mV offset to be 0.0 mV.