Hanna Instruments HI 3512 User Manual
Page 65

65
• electrode condition, with sign (3 chars). The
“-01” code means not calculated
• ISE calibration data (if available) which contains:
• the number of calibrated standards (1 char)
• the ISE unit (2 chars)
• the calibration slope, with sign and decimal point
(7 chars)
• the calibration time of the standard,
yymmdd
hhmmss (12 chars)
• standards information (for each standard)
• type (1 char): 0 - always standard solution
• status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration
• warnings during calibration (2 chars): 00 -
no warning
• standard value, with sign and decimal point
and exponent (12 chars)
• calibration time,
yymmddhhmmss (12 chars)
• Rel mV calibration data (if available), which contains:
• the calibration offset, with sign (7 chars)
• the calibration time,
yymmddhhmmss (12 chars)
Channel 2 GLP
• EC calibration data (if available), which contains:
• the number of calibrated standards (1 char)
• the offset, with sign and decimal point (7 chars)
• the cell constant, with sign and decimal point (7 chars)
• the calibration time,
yymmddhhmmss (12 chars)
• standards information (for each buffer)
• the unit of the standard (0 - microsiemens,
1 - milisiemens)
• the standard value, with sign, decimal
point and exponont (11 chars)
• the calibration time of the standard,
yymmddhhmmss (12 chars)
• the temperature reference value (2 chars)
• the temperature compensation mode (2 chars:
00 - NoTC, 01 - MTC, 02 - ATC)
• the temperature compensation coefficient, with