FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

• Immerse the pH electrode and the temperature

probe approximately 3 cm (1¼”) into a buffer
solution of your choice (pH 1.68, 4.01, 6.86,
7.01, 9.18, 10.01, 12.45 or a custom buffer)
and stir the buffer gently. The temperature probe
should be close to the pH electrode.

• Enter pH calibration. The instrument will display the measured pH,

the LCD first expected buffer and the temperature reading.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer,

Confirm

function key is displayed.

• Press

Confirm to confirm first point.

• The calibrated value and the second expected buffer value is then

displayed on the LCD.

• After the first calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 3 cm (1¾”) into the
second buffer solution and stir gently. The temperature probe should
be close to the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

3 cm