beautypg.com

Hanna Instruments HI 3512 User Manual

Page 67

background image

67

encoding
• 0x01 - beep ON (else OFF)
• 0x02 - degrees Celsius (else Fahrenheit)
• 0x04 and 0x08 - this 2 bits represent the

temperature compensation mode (10 - ATC,
01 - MTC, 00 - NoTC)

• The cell constant, with sign and decimal point (7 chars)
• The temperature compensation coefficient, with sign

and decimal point (6 chars)

• The temperature reference value (2 chars)
• The TDS factor, with sign and decimal point (5 chars)
• The short name of the selected language (3 chars)

NSLxy

Requests the number of logged samples (4 chars).
The command parameter (2 chars):

• x=1 request for channel 1
• x=2 request for channel 2
• y=P request for pH range
• y=M request for mV and Rel mV ranges
• y=I request for ISE range
• y=E request for EC range
• y=R request for Resistivity range
• y=T request for TDS ranges
• y=N request for NaCl range

LLSxy

Requests the number of lots (information about lots) on the specified
channel and range (x - channelnumber; y - range first letter)

• x=1 request for channel 1
• x=2 request for channel 2
• y=P request for pH range
• y=M request for mV and Rel mV ranges
• y=I request for ISE range
• y=E request for EC range
• y=R request for Resistivity range
• y=T request for TDS ranges
• y=N request for NaCl range

The answer string contains:

• number of lots (3 chars)

For each lot:

• the lot number (3 chars)
• the number of records in lot (3 chars)