• The calibrated value and the third expected standard value will be

displayed.

• After the second calibration point is confirmed, immerse the ISE

electrode approximately 2 cm (¾”) into the third calibration solution.

• If necessary, press the

ARROW keys to select a different standard

value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the

Confirm function key is displayed.

• Press

Confirm to confirm calibration.

• The calibrated value and the fourth expected standard value will be

displayed.

• After the third calibration point is confirmed, immerse the ISE electrode

approximately 2 cm (¾”) into the fourth calibration solution.

• If necessary, press the

ARROW keys to select a different standard

value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the

Confirm function key is displayed.

• Press

Confirm to confirm calibration.

• The calibrated value and the fifth expected standard value will be

displayed.

• After the fourth calibration point is confirmed, immerse the ISE

electrode approximately 2 cm (¾”) into the fifth calibration solution.

• If necessary, press the

ARROW keys to select a different standard

value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the

Confirm function key is displayed.

• Press

Confirm to confirm calibration. The instrument stores the

calibration value and returns to normal measurement mode.

Note: The instrument will automatically skip the standard solutions

used during calibration

.