Hanna Instruments HI 3512 User Manual
Page 30

30
FIVE-POINT CALIBRATION
Use part of this procedure for 2, 3 or 4 point
calibration. Press
ESC.
• Immerse the ISE electrode approximately 2 cm
(¾”) into the less concentrated standard
solution and stir gently.
• Press
CAL. The primary LCD will displays the ion concentration in the
selected unit or
“----” if not calibrated and first standard value.
• If necessary, press the
ARROW keys to select a different standard
value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the
Confirm function key is displayed.
• Press
Confirm to confirm calibration.
• The calibrated value and the second expected standard value will be
displayed.
• After the first calibration point is confirmed, immerse the ISE electrode
approximately 2 cm (¾”) into the second calibration solution.
• If necessary, press the
ARROW keys to select a different standard
value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the
Confirm function key is displayed.
• Press
Confirm to confirm calibration.
2 cm