Notes: • Buffers (standards) displayed in video inverse mode are

from previous calibrations (pH and ISE range only).

• The pH custom buffers are marked with an “

*” on the

right side of the buffer value.

LAST RELATIVE mV CALIBRATION DATA

Last Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press

GLP key while in Relative

mV measurement mode.
The instrument will display the Relative mV GLP information: calibration
date, time and offset.

LAST ISE CALIBRATION DATA

Last ISE calibration data is stored automatically after a successful
calibration.
To view the ISE calibration data, press

GLP function key while in ISE

measurement mode. If

GLP function key is not displayed press MENU

key.
The instrument will display the ISE calibration information: calibration
date, time, slope, calibration status and electrode type.