Hanna Instruments HI 3512 User Manual
Page 41

41
The following table lists the specific pH/ORP/ISE range parameters.
Item
Description
Valid value
Default
Calibribration
Number of days after Disable, 1 to 7 days
Disable
Timeout
Calibration warning
(pH/ISE)
is displayed
First point
Management of 1
Replace or offset
Replace
mode (pH)
point calibration
Custom buffer
Custom buffer
Max. 2 buffers
No
(pH)
setting
pH Resolution
Set pH resolution
0.1, 0.01
0.01
display
0.001
View calibr.
Display calibration
Enable or disabled
Enable
points (pH)
points
Display out of calibr.
Enable or disable
Enable
range warning
ISE probe
Type of ISE probe
Custom or Standard (17)
Fluoride
ISE unit
ppt, g/L, ppm, mg/L, ppb,
µ
g/L
ppm
mg/mL, M, mol/L, mmol/L, % W/V, user
Ion Charge
±1, ±2, none
+1
The following table lists the specific EC/TDS/NaCl/Resistivity range parameters:
Item
Description
Valid value
Default
Calibribration
Number of days after Disable, 1 to 7 days
Disable
Timeout
Calibration warning
(EC/Salinity)
is displayed
Out cal
Display warning if the
Enable/Disable
Disable
range check
reading is too far from the
(EC range only) calibration points
Temperature
No TC, MTC, ATC
ATC
compensation
mode
Range select
Fix one
Automatic, Fix one
Automatic
specific range
resolution of
EC, resistivity or TDS range
Cell constant
Manual set of
0.010 to 10.000
1.000
the cell constant
Temperature
Set the coeficient
0.00 to 10.00 %/°C
1.90 %/°C
coeficient
for linear temperature
compensation
Temperature
Reference temperature 15 °C, 20 °C, 25 °C
25 °C
Ref
Temperature unit
°C or °F
°C
TDS factor
0.40 to 1.00
0.50