• The unit of the second reading (1 char) (it is sent

only for Resistivity, TDS and Salinity ranges only). It
represents the unit of the EC reading (0 - microsiemens,
1 - milisiemens)

• Secondary reading (only when primary reading is

not mV or EC) - 8 ASCII chars, including sign and
decimal point.

• Temperature reading - 7 ASCII chars, with sign and

decimal point, always in ºC.

MDR

Requests the instrument model name and firmware code
(16 ASCII chars).

GLPx

Requests the calibration data record.

• x=1 calibration data from channel 1
• x=2 calibration data from channel 2

The answer string contains:

• GLP status (1 char): represents a 4 bit hexadecimal

encoding.
• 0x01 - pH calibration available (Ch1) or EC

calibration available (Ch2)

• 0x02 - Rel mV calibration available (Ch1) or

Salinity calibration available (Ch2)

• 0x04 - ISE calibration available

Channel 1 GLP

• pH calibration data (if available), which contains:

• the number of calibrated buffers (1 char)
• the offset, with sign and decimal point (7 chars)
• the average of slopes, with sign and decimal

point (7 chars)

• the calibration time,

yymmddhhmmss (12 chars)

• buffers information (for each buffer)

• type (1 char): 0 - standard, 1 - custom
• status (1 char): N (new) - calibrated in last

calibration; O (old) - from an old calibration.

• warnings during calibration (2 chars): 00 -

no warning, 04 - Clean Electrode warning

• buffer value, with sign and decimal point and

exponent (12 chars)

• the calibration time of the buffer,

yymmdd

hhmmss (12 chars).