Hanna Instruments HI 3512 User Manual
Page 24

24
Wrong old slope
An inconsistency between new and previous (old) calibration is detected.
Clear old calibration parameters and initiate calibration from the current
point. The instrument will keep all confirmed values during current
calibration.
Note: For single-point calibration the electrode condition is not displayed
in the measurement screen.
Each time a buffer is confirmed, the new calibration parameters replace
the older calibration parameters of the corresponding buffer.
If an additional single buffer calibration is added at a latter time, the
new buffer point will be added to the stored calibration.
If the existing stored calibration is full (five calibration points), after
confirming the calibration point, the instrument will ask which buffer will
be replaced by current buffer. On the Buffer line will be displayed the
proposed buffer.
Press
ARROW keys to select another buffer to be replaced.
Press
Confirm to confirm the buffer that will be replaced.
Press
ESC to exit mode. In this case, the buffer will not be entered.
Note: The replaced buffer is not removed from calibration list and it can
be selected for the next calibration points.
WORKING WITH CUSTOM BUFFERS
If at least one custom buffer was set in SETUP menu, it can be selected
for calibration by pressing the
ARROW keys. The Custom function key
will be displayed.
Press
Custom if you want to adjust the buffer to its value at the current
temperature.