:ROUTe[1..n]:SCAN:ADJust:AUTO
Description
This command sets the repeatability> option. When this option is selected, the switch mechanism returns to a reference position before aligning to a new position. This ensures optimum repeatability. When this option is not selected, the switch mechanism goes directly to the new channel position. This provides faster switching times, but less repeatability. On *RST, the value of this setting is OFF. Syntax :ROUTe[1..n]:SCAN:ADJust:AUTO Adjust> Parameter(s) AutoAdjust: defined as a The are accepted on input for increased readability. ON corresponds to 1 and OFF corresponds to 0. The disables the optimization for repeatability. 1 or ON, enables the optimization for repeatability. 0 or OFF, disables the optimization for repeatability.
The program data syntax for