Testing Fibers in Template Mode

68

FTB-7000 Series

Acquiring Traces in Template Mode

3c. Set the Template mode option you want to use on the current

trace acquisition:

³

Consider only the events already indicated on the reference trace
and ignore any other event occurring on the current trace.

³

Keep all the events on the current trace, whether they are on the
reference trace or not. You can delete these events later.

3d. Press Exit OTDR Setup to return to the main window.

3e. If you selected Create reference trace at step 4b, update your

reference trace as follows:

3f. Press Start or

from the keypad.

If the first connector check feature is enabled, a message will
appear if there is a problem with the injection level (see Enabling
or Disabling the First Connector Check
on page 28).

All traces will automatically be acquired and analyzed, and the
events will be identified.

I

MPORTANT

Once the first 15 acquisitions have been performed, or if you stop
the reference update manually, the only way to reactivate it will be
to close the application and start a new acquisition in Template
mode.