beautypg.com

EXFO FTB-7000 OTDR for FTB-200 v2 User Manual

Page 224

background image

Analyzing Traces with the Bidirectional Analysis Application

210

FTB-7000 Series

Viewing Results

Note: The attenuation value is always presented in dB per kilometers even if the

distance units you selected are not the kilometers. This follows the
standards of the fiber-optic industry that provides the attenuation values in
dB per kilometers.

Cumulative loss is calculated for the events displayed in the events
table. For the loss value of the complete link (fiber span), refer to the
loss measurement displayed in the Trace Info. tab.

If you want to modify events or fiber sections, see Modifying Events on
page 231,
Inserting Events on page 227, and Changing the Attenuation
of Fiber Sections
on page 237.

Event type detected
(see Description of Event Types on page 297)

Event number

Position: distance between the OTDR and the measured event, or
between the event and the beginning of the fiber span
OR
Length of a fiber section (distance between two events).

Reflectance measured at each reflective event
along the fiber

Attenuation (loss/distance) of individual

fiber section

Cumulative loss from span start to span end;
running total is provided at the end of each
event and fiber section.

Loss in dB for each event or fiber section

Unidirectional

traces

Injection level