Setting Pass/Fail Thresholds

The following table provides the default, minimum and maximum
thresholds.

Once the thresholds are set, the application will be able to perform
Pass/Fail to determine the status of the various events (pass or fail).

Values that are greater than the predefined fail thresholds are displayed in
white on a red background in the events table.

The Pass/Fail threshold LED, located on the front of the unit, will also
indicate the status (green for pass, red for fail).

Test

Default

Minimum

Maximum

Splice loss (dB)

1.000

0.015

5.000

Connector loss (dB)

1.000

0.015

5.000

Reflectance (dB)

–40.00

–80.0

0.0

Fiber section attenuation (dB/km)

0.400

0.000

5.000

Span loss (dB)

45.000

0.000

45.000

Span length (km)

0.0000

0.0000

300.0000

Span ORL (dB)

15.00

15.00

40.00