beautypg.com

EXFO FTB-7000 OTDR for FTB-200 v2 User Manual

Page 77

background image

Testing Fibers in Template Mode

OTDR

63

Restrictions of Template Mode

When you process traces using an OTDR, you acquire the traces as you go
along. When you process traces on a computer, you use traces stored on
disk; therefore, applying the span length is optional.

Item

To be valid...

Pulse width

³

Must be:

OR

Pulse width

³

This would also be valid:

OR

Fiber types

Compare singlemode traces with singlemode traces.
Compare multimode traces with multimode traces.

Number of

events

Traces must have at least two events (span start and span end) and a

fiber section.

Acquisition

mode

Reference trace must not be acquired in Real mode (see Setting Distance
Range, Pulse Width, and Acquisition Time
on page 45).

Wavelengths Reference wavelengths and wavelengths of subsequent (or reloaded)

traces must be identical.

Reference trace pulse

4

---------------------------------------------------

⎞ Current trace pulse

Reference trace pulse

Current trace pulse

Current trace pulse

Reference trace pulse

(

4

)

×

Reference trace pulse

Current trace pulse

Current trace pulse

4

------------------------------------------------

⎞ Reference trace pulse

Reference trace pulse

Current trace pulse

Reference trace pulse

Current trace pulse

(

4

)

×

Reference trace pulse

Current trace pulse