Testing Fibers in Advanced Mode

44

FTB-7000 Series

Setting the IOR, RBS Coefficient, and Helix Factor

2. From the OTDR Setup window, go to the Acquisition tab.

3. Use the up or down arrow located on the side of the wavelength box to

select the desired wavelength.

4. Select the default settings by pressing Default. When the application

prompts you, answer Yes only if you want to apply the new settings to
all wavelengths.

OR

Enter your own values in the boxes, for each available wavelength.

Note: You cannot define a different helix factor for each wavelength. This value

takes into account the difference between the length of the cable and the
length of the fiber inside the cable; it does not vary with wavelengths.

5. Press Exit OTDR Setup.

I

MPORTANT

Change the default RBS coefficient only if you have values provided
by the fiber manufacturer. If you set this parameter incorrectly, your
reflectance measurements will be inaccurate.

Wavelength for which

RBS and IOR will be

defined

Index of refraction

Rayleigh backscatter

coefficient