30

Calibration

5.5 Bipolar Output Calibration

If Bipolar mode for analog output is used, use Bipolar Output Calibration to
correctly calibrate. There are two steps involved: Gain Calibration and Offset
Calibration.

Gain Calibration:

1. Set

jumpers

JP1 and JP5 (for CH1) or JP2 and JP9 (for CH2) as

bipolar, and choose an internal reference as -5V.

2.

Set the digital data as (1000 0000 0000)

B

, trim VR4 (for CH1) or VR10

(for CH2) until the voltmeter reading is 0V.

Offset Calibration:

1.

Set the internal reference to -5V.

2.

Set the digital data as (0000 0000 0000)

B

, trim VR6 (for CH1) or VR12

(for CH2) until the voltmeter reading is -5V.

5.6 Current Sink Calibration

If using current sink output, please do the following:

Use -5V for internal reference and its reference voltage.

Follow the below procedures to calibrate the current sink output.

1.

Set the digital data as (0000 0000 0000)

B

, and internal reference as -

5V, using unipolar output range.

2.

Trim VR5 (for CH1) or VR11 (for CH2) until there is a constant current
of 4mA.