Remarks, Long press, Welcome – ThermoWorks 9882 PRINTING DATALOGGER THERMOMETER User Manual

Page 43

background image

MEAS

MEM

LOG

SET

Welcome

9851/ 9651 Hygro-Thermometer Calibration Procedure

EXIT

LXW

EDIT

SAVE

View Cali Data

Long press

EXIT

LXW

MANUAL

RH Cali Mode Select

INPUT

Zcal : -7340019 E-4
Scal : 5399497 E-5

RH 32.8 %

EXIT

Cali

PRN

Temp:

REC

RH

WBT

DP

60.8%

17.4C

20.0C

Temp:25.5C

Input probe cali data
Press F2
Calibrate manually
Press F3

ABORT

LXW

ENTER

SAVE

Input Cali Data

RH Manual Calibration

RH 60.8%

RH Manual Calibration

RH 75.3 %

RH Manual Calibration

ABORT

Remarks:

1. Be sure to keep a calibration
note enclosed in the package
This's for future RH probe cali-
bration use
2. While keyin the calibration value,

3. Press ABORT anytime to exit without save or press END to
save and exit if you don't want to wait for 30 minutes.
4. The LCD value will flash while doing manual 33% and 75%
calibration until the procedure finished.
5. Suggest to calibrate at stable room temp. environment.

Flashing for 30 mins

Flashing for 30 mins

No need to use
salt bottles

Need to use
salt bottles

(Need to plug probe into salt bottles)

41

Note:
"Input cali.data" method is for RH
probe replacement.
"Manual calibration" method is for
RH probe replacement & old probe
re-calibration.

Zcal :

E-4

Scal : 5399497 E-5

-7340019

EXIT

75.3

32.8

(Put probe into corresponding salt bottles
for 30 mins before press F2 or F3)

END

ABORT

END

END

refer to the value indicated on

the note. The values are slope and offset value of each probes.

Temp. 23.5 C

Temp. 23.5 C

Temp. 23.5 C