9812/9612 External Temp. Probe Calibration procedure

MEAS

MEM

LOG

SET

Welcome

C

EXIT

PRN

Press ON to Ir Meas

Ext:30.5
Ir:

CALI

Long press

EXIT

LXW

EDIT

SAVE

Probe Cali Data

Step 1 :0 at 5C
Step 2: -5 at 40C
Step 3: -6 at 85C
Note: Unit is 0.1C

ABORT

LXW

ENTER

SAVE

Input Cali Data

Step 1 :+

00

+/-

ABORT

LXW

ENTER

SAVE

Input Cali Data

Step 2 :+

00

+/-

Repeat until three points
value are saved

Remarks:

1. Be sure to keep a calibration
note enclosed in the package
This's for future probe calibration
use

2. While input the calibration value, refer to the value indicated
on the note. There are three values for different temperature.
The value is unique for each probe.
3. To exit manual calibration mode, long press F1 anytime to
exit.
4. +/- of the calibration means positive value or negative value.
5. Use 10 number keys of the meter to input the calibration value.