983X / 963X Manometer Calibration Procedure

ESC

LXW

ENTER

SAVE

Range: xx ~ xx psi

(remark2)

Long press

36

EXIT

Cali

PRN

Temp:

P: 0.1psi

ESC

LXW

ZERO

LOW

CALIBRATION

HI

REC

Cali:

PSI

(remak1)

* * *

Press up/down
or number keys to adjust

ESC

LXW

ENTER

SAVE

Cali:

PSI

(remak3)

* * *

Wait

Cali:

PSI

(remak3)

* * *

Range: xx ~ xx psi

Cali:

PSI

(remak4)

* * *

After 5 seconds

OK ( or NG)

EXIT

1.There are up to three calibration points. Please refer to below list:

Models

Cali.(HI)

psi range

Cali. (LOW)

Adjustable Range(psi)

9832/9632

9835/9635

98315/96315

9833/9633

9831/9631

0~+/-2psi

0~+/-5psi

0~+/-15psi

0~+/-30psi

0~+/-100psi

1.600psi

4.000psi

12.00psi

24.00psi

80.00psi

-1.600psi

-4.000psi

-12.00psi

-24.00psi

-80.00psi

1.400~1.800 / -1.400~-1.800

3.500~4.500 / -3.500~-4.500

10.50~13.50 / -10.50~-13.50

21.00~27.00 / -21.00~-27.00

70.00~90.00 / -70.00~-90.00