Welcome

9861/9661 pH/mV Probe Calibration procedure-1

1.PTS means the sensitivity of the probe, the acceptable slope
of electrode range is 85% to 105%.,please change the probe
to get correct reading.

Volt:

EXIT

CAL

PRN

Temp:

PH:

28C

7.60

-34.9mV

ATC

EXIT

LXW

ENTER

SAVE

Volt: -3.6mV
PH: 7.00

CALIBRATION

Auto Judging...
Last: 76 days ago

RST

CALIBRATION

PH:

7.00

EXIT

LXW

ENTER

SAVE

RST

Selecting point

ATC

CALIBRATION

PH:

7.00

EXIT

LXW

ENTER

SAVE

RST

Adj pH Value

ATC

CALIBRATION

Volt:-6.9mV

EXIT

LXW

ENTER

SAVE

RST

measuring....

ATC

Long press

To be continued..

Note This is to calibrate under
ATC( Temperature probe is
plugged side by pH probe )
To calibrate under NonATC,
please see procedure -3)

25

2.The meter is capable of calibrating up to 5 points using USA
or NIST pH buffer standard. There are 1.68-4-7-10-12.45 five
points for selection.

3.It is recommended that you start with first buffer at 7.00 pH .
The meter may select pH buffer value by pressing or
to calibrate with correct calibration value. +/-0.5 of the selected
range.