Remarks, 24 long press, Welcome – ThermoWorks 9882 PRINTING DATALOGGER THERMOMETER User Manual

Page 26

background image

EDIT

PRN

EXIT

NEXT

Set Clock:
MM-DD-YY HH:MM:SS

Set ID:

- - : :2 0

06 30 05 18 29

Disable

ID: User Name

EDIT

PRN

EXIT

NEXT

LCD Cont.(1-5):

Prn Cont. (1-9):

Unit:
Date:

Auto Off:

3

5

YY-MM-DD

20mins
Enable

Metric

EXIT

EDIT

BACK

PH
Volt
Temp

Select Item

PRN

MEAS

MEM

LOG

SET

Welcome

Volt:

EXIT

CAL

PRN

Temp:

PH:

28C

7.60

-34.9mV

ATC

EXIT

LXW

ENTER

SAVE

Volt: -3.6mV
PH: 7.00

"CALIBRATION"

Auto Judging...
Last: 76 days ago

RST

Remarks:

9861/9661 pH/mV Setting / Single measurement

1.No PRN function is existing for model 9661 meter.
2.Press CAL to enter calibration mode , "Auto judging...Last
xx days " means it has been xx days after the last calibra-
tion .
3.If you see a minus days "...- xx days " means you haven't
set the real time. Set real time from the main screen page.
4.Calibration is necessary and should be done regularly, re-
commend everyday if the meter is used often.
5.Meter features automatic buffer recognition to avoid errors.

24