CALIBRATION

PTS: 100.0%

TEMP: 27.0C

EXIT

LXW

ENTER

SAVE

RST

ATC

( Quit with save? )

Quit without save?

YES

NO

CALIBRATION

CALIBRATION

PTS:

75.0%

TEMP: 27.0C

EXIT

LXW

ENTER

SAVE

RST

ATC

Quit with save?

YES

NO

CALIBRATION

SAVE ENTER

CALIBRATION

Failure !!!

EXIT

LXW

RST

Ex. PTS's value is 75%

Remarks:

9861/9661 pH/mV Probe Calibration procedure-2

1.When you see "PTS:xx.x% , TEMP: xx.xC" shows on the
screen , that means the calibration of the point is done .
Press Enter to next calibration point or SAVE to stop cal
and store the calibration value .

2.You will see "Failure !!!" after you press YES to try to
save an unacceptable PTS range after calibration.
Under "Failure!!!" , press EXIT to exit calibration mode ,or
press RST to normal mode ( seeremark 3 for the details) .
Inactivate when pressing SAVE or ENTER after "Failure!!!" is
appeared.

3. RST stands for Reset , while proceeding any step, you
may stop calibration by pressing RST (F2) to back to
default setting.

Continuing from last page

Back to
Auto judging
next point cal.

Press EXIT ,"Quit without save?"
is appeared; Press SAVE,"Quit
with save ?" is appeared.

26