Caution – Oakworks Spine Positioning System User Manual

Page 5

background image

3

cautions

The cushioning foam contained within the Torso Support will lose its ability to

spring back to the original position over time and the amount of foam compression will

increase. Therefore, the Torso Support should be replaced periodically to ensure the

device functions as intended.

To prevent the potential of cross-contamination, it is strongly advised to use barrier tech-

niques when the device is in use. A disposable or laundered patient gown, or disposable pad

are satisfactory for use as a barrier for the Torso Support and other components and acces-

sories, except when the patient presents with pathology that would indicate otherwise. A

disposable face rest cover should be used to cover the Face Rest Cushion. Contact Oakworks

for ordering information. Barrier techniques should be used in addition to disinfection procedures,

not in lieu of them.

Be sure to support the weight of the patient’s head while making adjustments to the cervical

positioning feature of the Platform Frame. Make sure all cam locks are secure before

relinquishing support of the positioning assembly.

The Cervical Support System (see page 5) has metal parts that can cause back scatter of x-rays.

When x-rays are present, wear a suitable radiation barrier.

The Spine Positioning System is constructed using metal pins in the Quick Cam Locks and

aluminum tubing in the support structure. These are out of the field of view in most A-P and

oblique tilted views. Place the positioning assembly according to the recommendations in the

directions for use to eliminate, or reduce any artifacts. If artifacts still remain to the extent that

they would compromise the efficacy of needle placement, discontinue use of the device dur-

ing the affected procedure.

All materials used in the construction of the device and accessories are safe for

temporary and moderately frequent human contact. The device is not intended

for prolonged contact.

Do not use the Face Rest Support Arms as a handle to carry the Spine Positioning System.

Follow maintenance instructions found near the end of this manual. Mechanical components

should be checked periodically to insure that they are functioning properly to insure the safety

of the patient.

C

au

Tions

caution