maintenance:

Inspect Torso Support Pad monthly to be sure that the foam has not lost shape or

firmness to the extent that patient support would be compromised.

Inspect the base and components monthly to ensure that they have not been

damaged. Replace any damaged or worn components.

Inspect face rest platform locking mechanism weekly. Use the following procedure:

Step 1 - Lock the platform cam

Step 2 - Rock platform up & down

Gently rock platform up and down and note any “looseness” (some flexing

is normal). Look for gaps between the aluminum parts. If you feel “loose-

ness” or see gaps, cam tightening is required.

Inspect joints

Bad (gap)

Good (no gap)

face rest platform locking mechanism adjustment is