KERN & SOHN SFB 300K-1LM User Manual
Page 50
SFB-BA-e-1324
50
10 Data output RS 232C (optional)
Weighing data can be issued according to menu settings either via the RS 232C
interface or by pressing
via the interface.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing balance and the printer.
Use a suitable cable to connect the weighing balance to the interface of the
printer. Faultless operation requires an adequate KERN interface cable.
Communication parameters (baud rate, bits and parity) of balance and printer
must match. For detailed description of interface parameters see chap. 8,
menu block
„P1 COM“ or „P2 COM“.
10.1 Technical Data
10.2 Technical data
Connection
9 pin d-subminiature bushing
Pin 2 input
Pin 3 output
Pin 5 signal earth
Baud rate
Optional 600/1200/2400/4800/9600
Parity
8 bits, no parity / 7 bits, even parity / 7 bits, odd parity
10.3 Printer mode
Printout examples (KERN YKB-01N):
Weighing
ST, GS
1.000kg