KERN & SOHN SFB 300K-1LM User Manual
Page 24
SFB-BA-e-1324
24
6.6.2 Non verifiable models
Call up menu:
1. Switch-on balance and during the selftest press
.
2. Press subsequently
,
,
the first menu block
„PO CHK“ will be displayed.
3. Press
repeatedly until „P3 CAL“ will be displayed.
4. Confirm with
; press
repeatedly until „CAL“
appears.
5. Acknowledge using
, the current setting is displayed.
Press
to confirm; press
to select setting.
noLin = adjustment
LineAr = linearization, see chap. 6.6
How to carry out adjustment:
Confirm menu setting „noLin“ by
.
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press
.
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with
.
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press
.
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically. An adjusting error or
incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.