KERN & SOHN SFB 300K-1LM User Manual
Page 32
SFB-BA-e-1324
32
Press the navigation keys (See chap. 2.1.1) to enter the
upper limit, e.g. 1,100 kg; the currently enabled digit will be
flashing.
Confirm input by
.
Press
repeatedly until
is displayed.
Press
and the current setting for the acoustic signal will
be shown.
Select desired setting (no, ok, ng) by
.
Confirm input by
.
Press
; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be
weighed are within the two tolerance limits.
Weighing with tolerance range
Tare when using a weighing container.
Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified
tolerance
Load within specified
tolerance
Load exceeds specified
tolerance
Red signal light
next to „-“ ON illuminated
Green signal light
next to „„ illuminated
Red signal light
next to „+“ ON illuminated
The tolerance control is not active when the weight is under 20d.
To delete limits, enter “00.000 kg“.