Remove the transportation lock:
1. Models platform size 300 x 240 mm
Remove the marked screws.
2. Models platform size 400 x 300 mm
Remove the screw marked by the label
.
Attention: The sealed screws must not be unscrewed.
Accurate weighing results require a weighing bridge with perfect horizontal
alignment. During initial installation and after each change of work area it is
necessary to level the weighing bridge.
As the air bubble is located under
the weighing plate, remove it.
Level balance with foot screws until
the air bubble of the water balance
is in the prescribed circle.
This manual is related to the following products: