KERN & SOHN SFB 300K-1LM User Manual
Page 45
SFB-BA-e-1324
45
8.2 Overview verified models
In verified weighing systems the access to „P2 mode and „P4 tAr“ is locked.
In order to unlock the access, the seal must be destroyed and both contacts of the
printed circuit board [K2] must be short-circuited by a jumper, see chap. 6.11.
Attention:
After destruction of the seal the weighing system must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can be
reused for applications subject to verification.
Menu block
Main menu
Menu item
Submenu
Available settings / explanation
PO CHK
Weighing with
tolerance range,
see chap. 7.7
SET H
Upper limit value „Tolerance check weighing“, input see chap. 7.7.1
SET LO
Lower limit value „Tolerance check weighing“, input see chap. 7.7.1
PCS H
Upper limit value „Tolerance check counting“, input see chap. 7.7.2
PCS L
Lower limit value „Tolerance check counting“, input see chap. 7.7.2
BEEP
no
Acoustic signal for weighing with tolerance
range switched off
ok
Audio sound when load is within tolerance
limits
ng
Audio sound when load is beyond tolerance
limits
P1 COM
Interface
parameter
MODE
CONT
Continuous data output
ST1
One output for stable weighing value
STC
Continuous data output of stable weighing
values
PR1
Output after pressing
PR2
Manual totalizing, see chap. 7.8
Press
and the weighing value will be
added to the summation memory and issued.
AUTO
For automatic totalizing see chap. 7.9. This
function is used to issue and add individual
weighing values automatically to the
summation memory on unloading of weighing
scale.
ASK
For remote control commands, see chap. 10.4
wireless
Not documented
Kit 1
baud
Available Baudrate: 600, 1200, 2400, 4800, 9600
Pr
7E1
7 bits, even parity
7o1
7 bits, odd parity
8n1
8 bits, no parity
PtYPE
tPUP
Standard printer setting
LP50
Not documented
Lab
Lab x
Details see following table 1
Prt
Prt x