35
7.8 Manual totalizing
With this function the individual weighing values are added into the summation
memory by pressing
and edited, when an optional printer is connected.
Menu settings:
„P1 COM“ or „P2 COM“ „MODE“ „PR2““, see chap. 8
The totalisation function is not active when the weight is under 20d.
Add up:
Place goods to be weighed A.
Wait until the stability display
STABLE
appears, then press
. The weight value
will be saved and a printout received if an optional printer is connected.
Remove the weighed good. More weighed goods can only be added when the
display = zero.
Place goods to be weighed B.
Wait until the stability display appears, then press
. The weight value will be
added to the summation memory and possibly printed. The number of weighing
processes followed by the total weight will be shown for 2 sec.
Add more weighed goods as described before.
Please note that the weighing system must be unloaded between the individual
weighing procedures.
This process may be repeated 99 times or till such time as the capacity of the
weighing system has been exhausted.
Display and output sum „Total“:
Press
and the number of weighings followed by the total weight will be
shown for 2 sec. To receive a printout, press
during this display.