KERN & SOHN SFB 300K-1LM User Manual
Page 26
SFB-BA-e-1324
26
6.7.2
Non-verified models
Call-up menu item P3 CALCalLiner, see chap. 6.5.1
Confirm by
, the password query „Pn“ will be displayed.
Press subsequently
,
,
or
,
,
.
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press
.
When “Ld 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait
for stability display, then press
.
When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press
.
When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait
for stability display, then press
.
After a successful linearisation the balance will carry out a
self-test. Remove adjusting weight during selftest, balance
will return into weighing mode automatically.