5.
SET UP AND CALIBRATION
The chapter refers to both the FDT and the FAD setting up.
The FAD is configured with Fn 49 procedure. The RTD–52 is calibrated with Fn 46 procedure.
For setup and calibration of the FAD the unit must have been connected and the correct
communication protocol installed in both the FAD and FDT.
It is advisable to leave the default settings [ (9600, E, 7, 1) , address A].
If there is no communication, perform Fn 33 to “auto tune” the FDT to the FAD communication speed
and address code. In the setup dialogue the following two numeric scroll edit procedures are used.
1.
SCROLL
2.
NUMSCROLL
EDIT
I
= DISPLAY NEXT SELECTION
T= SELECT DIGIT (CYCLIC)
P = ACCEPT DISPLAYED VALUE
I
= INCREMENT FLASHING DIGIT
P
= ACCEPT DISPLAYED NUMBER
5.1 FDT CONFIGURATION
To enter the configuration procedure of FDT press a . The display shows Fn 00.
Use “NUMSCROLL EDIT” until the display shows Fn 46, then press P .
Display shows SETUP . Press I to step through SETUP, STORE, INIT dialogues.
Press O key to exit the menu or press P to enter the displayed item.
SETUP SUBMENU.
Press I to scroll through SETUP 1, SETUP 2, SETUP 3, SETUP.
Press P to enter the displayed dialogue.
SETUP 1
(OPERATIONAL PARAMETERS).
1.1
Totalizing memory
0 = No
1 = Yes
Enables the totalizing of printed
weights.
1.2
Alibi memory
0 = No
1 = Yes
enables the electronic tally roll.
(B only)(see chapter 4.2 & 4.4)
1.3 Reserved
1.4
Date format
0 = d-m-y
1 = m-d-y
1.5 Reserved
1.6
Real time clock
0 = No
1 = Yes
&
ram
totaliser
Enables
date
time
and
totalizing
in
battery backed ram. (B only)
1.7 Reserved
1.8 Reserved
1.t
Display off timer
00-99 minutes 00=Disabled.
Time for the display to turn off
automatically if the weight is
zero and no key is pressed.