FLINTEC
Document
Date
Code
Rev.
Page
FDT – VERSION A/B – TECH. -REFERENCE MANUAL
05.04.99
FDT.RM
2
17
of 23
Fn 42
DOWNLOAD CUSTOM PRINT FORMAT . 2
The display shows For 2 indicating that it is ready to receive format 2.
Begin the transmission. When the transmission ends press P . the display shows Prn 2
briefly when the format is saved in non volatile memory. A max of 512 bytes may be stored.
If more characters are received they will be lost.
Fn 46
SETUP OF FDT
(Refer to chapter 5)
Fn 48
VIEW CALIBRATION AUDIT TRAIL COUNTER
SEALEd is displayed if the CAL LOCK jumper inside the FAD is inserted. Press P to continue.
CALPin is displayed if the software PIN calibration lock has been activated. Press P to continue.
The 6 digit AUDIT TRAIL counter is displayed. Press to P to exit.
The audit trail counter is incremented every time a weight parameter or calibration is
changed regardless if the change was saved in EEPROM or not.
Fn 49
SETUP & CALIBRATION OF FAD (Refer to chapter 5)
Fn 55
VIEW ALIBI MEMORY LOCATION / PRINT 10 NEXT SERIAL NUMBERS
The display shows n 1 2 3 4 where 1234 is the serial number of the last record.
Key in the 4 digit serial number desired and / or press P . The display flashes the weight
of this record. Press P to print the serial number displayed plus the next nine locations or
press
O to exit.
Fn 56
PRINT ALL ALIBI MEMORY
Automatically prints a list of the contents of the Alibi memory. The printer must be
capable of printing on 80 columns paper compressed mode.
Empty locations are printed - - - - - - .
Corrupted locations are printed * * * * * * .
The program exits when the printout is completed or O pressed.
Fn 57
CHECK ALIBI MEMORY
A checksum is performed on each Alibi memory record. If an error is found E77 is displayed.
Press
O to exit. If all is OK PASS is displayed briefly.
Fn 80
LOADCELL MV METER
The actual mV/V output of the scale sensors is displayed.
To act as a mV/V meter the unit loads default calibration data.
Press O to end. The unit will restart.
Fn 81
DISPLAY INTERNAL A/D COUNTS
The display shows Corn 0 briefly. Then it shows the sum of the corrected loadcell
output(s) in internal resolution. Press I .
The display shows Corn 1 briefly, then loadcell 1 output raw A/D. Press I .
The display shows Corn 2 briefly, then loadcell 2 output raw A/D. Press I .
The display shows Corn 3 briefly, then loadcell 3 output raw A/D. Press I .
The display shows Corn 4 briefly, then loadcell 4 output raw A/D. Press O to exit.
Fn 82
DISPLAY CORNER FACTORS
Displays the calculated corner factors for the connected loadcells.
The display shows FAct 1 briefly, then loadcell 1 correction factor. Press I.
The display shows FAct 2 briefly, then loadcell 2 correction factor. Press I.
The display shows FAct 3 briefly, then loadcell 3 correction factor. Press I.
The display shows FAct 4 briefly, then loadcell 4 correction factor. Press O to exit.
Fn 99 SOFT RESET FAD (Similar to FAD power on).