FDT – VERSION A/B – TECH. -REFERENCE MANUAL
05.04.99
FDT.RM
2
14
of 23
E-CAL DIALOGUE
(Electronic calibration of FAD with mV/V from loadcell data sheets)
Display shows CORNER. Press P to calibrate corners.
Display shows 000000 . Enter the rated output (in mV/V) of loadcell in corner 1.
(Example : If R. O of Loadcell 1= 1,9793 mV/V then enter 019793).
The display shows CORNER.2 . Press P . Enter the rated output (in mV/V) of loadcell in corner 2.
The display shows CORNER.3 . Press P . Enter the rated output (in mV/V) of loadcell in corner 3.
The display shows CORNER.4 . Press P . Enter the rated output (in mV/V) of loadcell in corner 4.
Display shows WAIT… for about 5 sec. Then it shows weight.
Press O . Display shows CORNER. Press I . Display shows ZERO.
Press P . Enter the overall mV/V of the deadload. (Loadcell zero balance + scale deadload).
Display shows WAIT… and then it displays weight.
Press O then I .
The display shows SPAN. Press P .
The display shows C xxxxx . Enter the rated capacity of each loadcell.
The display shows WAIT… and then it displays weight.
Press O , O , I to exit .
SET UP SUBMENU
Press I to scroll through SET UP 1, SET UP 2, SET UP.
Press P to enter the display dialogue.
SET UP 1 (operational parameters of FAD)
1.c = Tilt switch
0 = NO
1 = N.C
1.t = Debounce time
0 – 9 sec
If both 1.c = 0 and 1.t = 0 the tilt function is disabled.
SET UP 2 (COMMUNICATION PARAMETERS OF FAD)
Display shows SEt UP2 .
Press
P
to enter the dialogue.
2.b
BAUD RATE
24 = 2400 baud
38 = 38.4kbaud
Use I to select
96 = 9600 baud
57 = 57.6kbaud
and
P to accept
19 = 19200 baud
11 = 115kbaud
and
proceed.
2.d
DATA BITS & PARITY
08 = 8 data bits, no parity
Use I to select
17 = 7 data bits, Even parity
and
P to accept
2.A
FAD ADDRESS
Use “DIGIT ENTRY” to key in the FAD network address
(65
–
127)
(decimal).
DIALOGUE
STORE
(Save in EEPROM)
Press
P
to permanently save all the calibration data.
Program exits setup / calibration and reinitialises. (up to 100.000 storage cycles allowed).
The data will be stored in the FAD EEPROM.