Flintec – Flintec FDT-B User Manual

Page 14

background image

FLINTEC

Document

Date

Code

Rev.

Page

FDT – VERSION A/B – TECH. -REFERENCE MANUAL

05.04.99

FDT.RM

2

14

of 23


E-CAL DIALOGUE

(Electronic calibration of FAD with mV/V from loadcell data sheets)

Display shows CORNER. Press P to calibrate corners.
Display shows 000000 . Enter the rated output (in mV/V) of loadcell in corner 1.
(Example : If R. O of Loadcell 1= 1,9793 mV/V then enter 019793).
The display shows CORNER.2 . Press P . Enter the rated output (in mV/V) of loadcell in corner 2.

The display shows CORNER.3 . Press P . Enter the rated output (in mV/V) of loadcell in corner 3.

The display shows CORNER.4 . Press P . Enter the rated output (in mV/V) of loadcell in corner 4.
Display shows WAIT… for about 5 sec. Then it shows weight.
Press O . Display shows CORNER. Press I . Display shows ZERO.

Press P . Enter the overall mV/V of the deadload. (Loadcell zero balance + scale deadload).
Display shows WAIT… and then it displays weight.
Press O then I .

The display shows SPAN. Press P .
The display shows C xxxxx . Enter the rated capacity of each loadcell.
The display shows WAIT… and then it displays weight.
Press O , O , I to exit .


SET UP SUBMENU

Press I to scroll through SET UP 1, SET UP 2, SET UP.

Press P to enter the display dialogue.

SET UP 1 (operational parameters of FAD)

1.c = Tilt switch

0 = NO

1 = N.C

1.t = Debounce time

0 – 9 sec

If both 1.c = 0 and 1.t = 0 the tilt function is disabled.

SET UP 2 (COMMUNICATION PARAMETERS OF FAD)

Display shows SEt UP2 .

Press

P

to enter the dialogue.

2.b

BAUD RATE

24 = 2400 baud

38 = 38.4kbaud

Use I to select

96 = 9600 baud

57 = 57.6kbaud

and

P to accept

19 = 19200 baud

11 = 115kbaud

and

proceed.

2.d

DATA BITS & PARITY

08 = 8 data bits, no parity

Use I to select

17 = 7 data bits, Even parity

and

P to accept

2.A

FAD ADDRESS

Use “DIGIT ENTRY” to key in the FAD network address

(65

127)

(decimal).

DIALOGUE

STORE

(Save in EEPROM)

Press

P

to permanently save all the calibration data.

Program exits setup / calibration and reinitialises. (up to 100.000 storage cycles allowed).
The data will be stored in the FAD EEPROM.