beautypg.com

Flintec, Communications – Flintec FDT-B User Manual

Page 19

background image

FLINTEC

Document

Date

Code

Rev.

Page

FDT – VERSION A/B – TECH. -REFERENCE MANUAL

05.04.99

FDT.RM

2

19

of 23

8. COMMUNICATIONS

Version A has one communication port RS485A (SERIAL PORT 1). This is the port used for

connection to FAD(s) and for networking.

Version B has in addition to serial port 1 an RS232C port (SERIAL PORT 2), used for connection

with printers or personal computers.


8.1

SERIAL PORT 1 (RS485A)

The port operates in half duplex mode. Communication is based on “LESIP” master / slave serial

interface protocol. (See “LESIP” with FAD appendix for details).

Baudrate and databits are configured in FDT setup 3.

Cable connections are shown in Annex A “CONNECTOR & CABLE DIAGRAM”.



8.2

SERIAL PORT 2 (RS232C , VERSION B ONLY)

The port operates if full duplex mode. Baudrate is fixed at 1200 bits/sec. 7 data bits, even parity or

8 data bits no parity may be selected in FDT SET UP 2.d. The functionality of the port is

also programmed in set up 2. The different modes of operation are detailed below:


8.2.1

CONTINUOUS WEIGHT OUTPUT
The indicated weight and status information is transmitted continuously. No handshake is required.
Data is output from Pin 3 = Tx and Pin 5 = GND

SET UP

REQUIREMENTS (1200, E, 7, 1)

1 Don’t

care

2

2.t = 02, 2.d = 17 the rest don’t care

3

According to application

DATA BLOCK COMPOSITION ( P +123.45 CR )

BYTE

NAME


DESCRIPTION

1







2

3-8

9

WEIGHT

STATUS






POLARITY

WEIGHT

DIGITS

SYNC


Bit0

0=NORMAL 1=NO WEIGHT DISPLAY

Bit1

0=GROSS

1=NET

Bit2

0=

1=AUTO ZERO

Bit3

0=WITHIN RANGE 1=OUT OF RANGE

Bit4

0=NO STANDSTILL 1=STANDSTILL

Bit5

0=NORMAL

1=UNDER MIN. WEIGHING RANGE

Bit6

ALWAYS 1 TO OBTAIN PRINTABLE CHARACTERS

Bit7

ZERO OR PARITY

"+" OR "-"
6 DIGITS DEPENDING ON D 500 N SET UP
INCLUDING DECIMAL POINT IF ANY
CR (0d hex) FOR SYNCHRONISATION