7. Maintenance and Troubleshooting
A
CAUTION
■ Do not conduct wiring work when
■ Do not touch the LF470 main body
power is applied.
when high temperature fluid is be-
ing measured.
Wiring while power is
The fluid raises the main
\^Z applied can cause electric
V^/ body temperature and can
DON’T
DON’T
cause burns.
7.1 Maintenance
IMPORTANT
It is recommended that the detector pipe be cleaned once a year.
■ Cleaning the pipe wall inside of detector
If the fluid to be measured contains slurry; a high concentration of electrically conductive
solids, the slurry may accumulate as sticking on the pipe wall inside of detector.
The sticking causes a reduction of flow measuring outputs when it is not cleaned inside pipe
of detector for a long time.
The flowmeter needs to make sure it remains around the wall inside pipe or not when the
following condition had happened.
(1) The flowmeter becomes a reduction of its output.
(2) The flowmeter caimot improve its output in spite of being calibrated.
The flowmeter is required to remove along with clean it using a soft brush. It will be come
back the usual and stable flow measurement after cleaning.
The pipe wall is never scratched when cleaning.
*Toshiba recommends to clean inside pipe periodically when this condition will be
come easily.
*It is better to choose the suitable diameter which becomes more than 3 m/s as its
flow velocity to avoid this condition generally.
Notes: Before uninstalling the flowmeter(detector) from the pipe, necessarily confirm the
fluid of temperature and chemical property, and empty the fluid from detector. Pre
pare the way to protect operator and system from bad influence; high temperature
fluid and chemical property
- 2 6 -