beautypg.com

2 external device connections and grounding, Important, Toshiba – Toshiba LF470 User Manual

Page 21

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

background image

TOSHIBA

6,F^8A0,8,0,2

5.2 External Device Connections and Grounding

The connections for the LF470 detector are shown in Figure 5.1 and 5.2.

Proceed with wiring as described in Section 5.3, “Wiring Procedure.”

IMPORTANT

(1) The input signal cable and excitation cable are attached to the detector. Be sure to use the at­

tached cable, check the specification of the detector.

(2) The converter side ends of the attached cables are covered with caps to prevent exposure to

humidity. Do not remove the caps from the cable until directly before wiring to the converter.

(3) Be sure to use thick walled steel conduit (22mm) for each the signal cable and the excitation

cable between the detector and the converter, and locate it as far as possible from other great

current cabling and it is not parallel with such cabling. Use flexible conduits at the cable outlets

of the detector.

(4) The FG terminal of the converter should be grounded with 100 ohm or less ground resistance.

Use a heavy copper braid or wire (cross-sectional area 5.5 mm^ minimum) to ground the termi­

nal and make it as short as possible.

for Converter

/■—^\ r-S

Excitation cable

(Three-core chloroprene cabtyre cable)

- 2 0 -