(3) Flow Direction
Install the detector in accordance with the flow direction arrow on the detector. See Figure 4.4.
over
side
Figure 4.4 Flow direction arrow on the detector
(4) Preventing an Empty Pipe Condition
Design an upright pipe run (Figure 4.5) or sufficient head pressure (Fig. 4.6) at the
downstream detector outlet if there is a possibility of the detector pipe becoming emptied.
Upright pipe run
Figure 4.5 Detector with an upright pipe run at downstream outlet
Detector
/
Figure 4.6 Detector with sufficient head pressure at downstream outlet