Normalization performed by mistake on calibration
standard instead of on uncoated specimen.

Perform normalization on uncoated specimen.

Er14

Internal instrument error: Not able to compute
master characteristic. The original master
characteristic is retained.

Repeat master calibration.
If this error occurs repeatedly:
Contact customer service.

Er15

Not able to save the master characteristic.
Probe may be defective.

Contact customer service.

Er17

Not enough readings. Thus, not able to finish a
block (applies to measurements w. fixed block
limits).

Make additional measurements until the first
block can be finished. Then generate a new final
result.

Er22

Internal instrument error. Error accessing the
measurement memory or. EEPROM.

Contact customer service.