14 Trouble Shooting

Message

Explanation / Possible Cause

Solution

Er1

Internal error

Perform a normalization. If the error occurs
repeatedly, call the Fischer service dpt.

Er4

Overflow of the data memory

Delete the data memory.

Er5

Cannot determine value for substrate material.

Perform calibration on coating using a suitable
calibration standard.
The selected calibration standard should have at
least 50% of the coating thickness of the coated
specimen (this refers to the specimen that is
used for the calibration on the coating).