beautypg.com

Calibration test report using an external weight – A&D Weighing HR-202i User Manual

Page 40

background image

38

Calibration test report using an external weight

Note Calibration test does not perform calibration.


Key operation
1 Press and hold the

CAL

key to display

CC out

and release the key.

2

Cal 0

is displayed.

3 When updating the target value, press the

RANGE

key and proceed to step 4.

When using preset target value, proceed to step 5.

4 Specify calibration mass value using the following keys.

RANGE

key ........ The key to select the blinking figure

RE-ZERO

(+)key .. The key to increase the value of the blinking figure.

MODE

(-)key........ The key to decrease the value of the blinking figure.

PRINT

key ......... The key to store the new value.

5 When pressing the

RE-ZERO

key, the zero point is measured and the weight value is

displayed for a few seconds.

6 Place the displayed mass on the pan. Press the

PRINT

key to measure the mass and the

weight value is displayed for a few seconds.

7 If GLP output is used,

glp

is displayed and the calibration test report is output.

8 The balance returns to the weighing mode after this test.


Setting of "info 1"

Setting of "info 2"

AD-8121 printer format

General format



Space, ASCII 20h

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration test type

Zero point value

Target weight value

Target weight

Signature

-----------

A

-

&

-

D

MODEL

----

HR-300i

S/N

-----

01234567

ID

------

LAB-0123

DATE

--

2004/09/30

TIME

----

12:34:56

CAL.TEST(EXT.)
ACTUAL

-------

0.0000

--

g

----

+200.0002

--

g

TARGET

----

+200.0000

--

g

SIGNATURE


-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

The space

to fill date in.

The space

to fill time in.

This manual is related to the following products: