beautypg.com

Glp report, Calibration report using an external weight – A&D Weighing HR-202i User Manual

Page 39

background image

37

11.2.

GLP Report

Set the following parameters to output the report.

Refer to "13.2. Connection to peripheral equipment", for connection to an AD-8121B.
To print the report, set the "GLP output (

info

)" parameter to "

1

" and use MODE 3 of the

AD-8121B. If MODE1 is used, select temporary dump print mode by pressing the

STAT.

key of the AD-8121B.

To output the report to a personal computer using the RS-232C interface, set the "GLP

output (

info

)" parameter to "

2

".

If the time and date are not correct, adjust the settings of the AD-8121B.

Calibration report using an external weight

Key operation
1 Press and hold the

CAL

key to display

Cal out

and release the key.

2

Cal 0

is displayed.

3 When updating the calibration mass value, press the

RANGE

key and proceed to step 4.

When using the preset calibration mass value, proceed to step 5.

4 Specify calibration mass value using the following keys.

RANGE

key ........ The key to select the blinking figure

RE-ZERO

(+)key .. The key to increase the value of the blinking figure.

MODE

(-)key........ The key to decrease the value of the blinking figure.

PRINT

key ......... The key to store the new value.

5 When pressing the

RE-ZERO

key, the zero point is measured and the weight value is

displayed for a few seconds.

6 Place the displayed mass on the pan. Press the

PRINT

key to measure the mass and the

weight value is displayed for a few seconds.

7 If GLP output is used,

glp

is displayed and the calibration report is output.

8 The balance returns to the weighing mode after this calibration.

Setting of "info 1"

Setting of "info 2"

AD-8121 printer format

General format


Space, ASCII 20h

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration type

Calibration weight

Signature

-----------

A

-

&

-

D

MODEL

----

HR-300i

S/N

-----

01234567

ID

------

LAB-0123

DATE

--

2004/09/30

TIME

----

12:34:56

CALIBRATED(EXT.)
CAL.WEIGHT

----

+200.0000

--

g

SIGNATURE


-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

The space

to fill date in.

The space

to fill time in.

This manual is related to the following products: