beautypg.com

Response adjustment – A&D Weighing HR-202i User Manual

Page 21

background image

19

7. Response

Adjustment

This function adjusts the response of the balance and reduces the influence of drafts and/or
vibration at the place where the balance is installed. The function has three stages as
follows :

Indicator Parameter Response

Stability

FAST

Cond 0

MID.

Cond 1

SLOW

Cond 2

Fast response, Sensitive value

Slow response, Stable value



1 Press and hold the

MODE

key until

RESPONSE

is

displayed, then release the key.
Press the

MODE

key again quickly.



2 Select a stage of the response adjustment using

the

MODE

key. Either

FAST

,

MID.

or

SLOW

can be selected.




3 The balance automatically returns to the weighing

mode and displays the updated response indicator
for about thirty seconds.




Advise

If the automatic response adjustment is not helpful, specify a parameter for "Condition (

Cond

) "

of "Environment, Display (

ba5fnc

)" with key operation.



.

Response

indicator

Release and wait

Press and hold

Release

Press again

immediately

Select a parameter

This manual is related to the following products: