Initializing the balance – A&D Weighing HR-202i User Manual
Page 26

24
9.2.
Initializing the Balance
This function returns the following parameters to factory settings.
Calibration
data
Function
table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
External calibration weight and target mass value
Function switch settings ("9.1. Permit or Inhibit")
Note Be sure to calibrate the balance after initialization.
1 Press
the
ON:OFF
key to turn off the display.
2 While pressing and holding the
key and the
RANGE
key, press the
ON:OFF
key to display
p5
.
3 Press
the
RANGE
key to display
Clr
.
4 Press
the
key.
To cancel this operation, press the
CAL
key.
5 Press
the
RE-ZERO
key to display
Clr go
.
6 Press
the
key to initialize the balance.
The balance will automatically return to the weighing mode.
Press and hold Press