15

6.2.

Dual Range

HR-202i

Weighing range

Available minimum display

Precision range

0 g to 51 g

0.01 mg

0.1 mg

1 mg

Standard range

51 g to 220 g

0.1 mg

1 mg


The

HR-202i is equipped with two

ranges, "precision range" and
"standard range".

When weighing is started by

pressing the

ON:OFF

key, the

minimum display will be 0.1 mg.

Pressing the

RANGE

key will

switch these ranges alternately.




When a sample is weighed using the

precision range with a minimum display of
0.01 mg and the weight value exceeds the
precision range value, the minimum display
changes to 0.1 mg of the standard range.
When removing the sample, the minimum
display changes to 0.01 mg of the precision
range automatically.



When a tare weight (container mass value)

exceeds the precision range value, even if
the sample is within the precision range
value, 0.01 mg of the precision range can
not be selected for the minimum display. In
order to select 0.01 mg of the precision
range, remove the tare weight and press
the

RE-ZERO

key to cancel it.



When a minimum display of 0.1 mg or 1 mg is

selected by the

RANGE

key, the minimum

display is maintained even if the range is
changed during weighing.

Turn on the balance

Turn on the balance

Minimum display

0.1 mg

Minimum display

0.01 mg

Minimum display

1 mg

Precision Range

Minimum display

0.1 mg

Minimum display

1 mg

Standard Range

Minimum display

0.01 mg

Minimum display

0.1 mg

Minimum display

0.01 mg

Minimum display

0.01 mg

Minimum display

0.1 mg

Zero display after

removing the container

Minimum display 0.01 mg

Zero display after

placing a container

Minimum display 0.1 mg